RadFET dose response in the CHARM mixed-field: FLUKA MC simulations
CERN, European Organization for Nuclear Research, Geneva, Switzerland
⁎ e-mail: matteo.marzo@cern.ch
Received:
18
January
2017
Received in final form:
16
May
2017
Accepted:
19
June
2017
Published online: 28 July 2017
This paper focuses on Monte Carlo simulations aimed at calculating the dose response of the RadFET dosimeter, when exposed to the complex CHARM mixed-fields, at CERN. We study how the dose deposited in the gate oxide (SiO2) of the RadFET is affected by the energy threshold variation in the Monte Carlo simulations as well as the materials and sizes of scoring volumes. Also the characteristics of the input spectra will be taken into account and their impact on the final simulated dose will be studied. Dose variation as a function of the position of the RadFET in the test facility will be then examined and comparisons with experimental results will be shown. The contribution to the total dose due to all particles of the mixed-field, under different target-shielding configurations, is finally presented, aiming at a complete characterization of the RadFETs dose response in the CHARM mixed-fields.
© M. Marzo et al., published by EDP Sciences, 2017
This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
1 Introduction
We present Monte Carlo calculations of the dose response of the RADiation sensing Field Effect Transistors (RadFET) [1–4], when they are exposed to the radiation environment at the CERN High Energy Accelerator Mixed-field (CHARM) test facility [5,6].
The main purpose of the facility is to replicate different radiation environments (space, atmospheric, accelerator complexes, for instance) for radiation effects testing on electronic components and systems. RadFETs at CHARM are then used as online dosimeters during the radiation tests, to characterize the complex mixed-fields reproduced in the test area.
In this regard, we are interested in investigating the dose deposited in the gate oxide (GO) of the RadFET. The GO is a very small sensitive volume where the increase of the accumulated charge δQtot, as a consequence of the exposure to the radiation fields of interest, implies a change of the source–drain channel thickness of the MOSFET. This causes a drift in the threshold voltage δVth of the device and a modification of its electric response, which can therefore be correlated to the deposited dose.
Measurements and FLUKA [7,8] simulations of the dose response of RadFETs in the mixed-field at CHARM had been already performed to understand the radiation field in the test area. However discrepancies in the 25–45% range had been identified between simulated and experimental values. In this paper we want to set up more accurate simulations to investigate this mismatch.
2 CHARM and its mixed-field: measurements vs. FLUKA simulations
The CHARM test facility is located in the Meyrin site of CERN, specifically in the Proton Synchrotron (PS) East Area hall. The main source of the radiation field reproduced at CHARM is in fact the particle shower developing from the interaction between the 24 GeV/c proton beam extracted from the PS and a metallic target [5,6].
2.1 Facility variables
Due to the high energetic particles treated at CHARM, particular attention is given to the protection of the personnel. Concrete, marble and iron are used to separate the irradiation chamber from the technical area. The irradiation chamber is a 7 × 7 × 3 m3 room, available for the irradiation tests.
The most relevant variables needed to obtain and tune the facility's mixed-field, are:
the 24 GeV/c beam from the PS;
a target placed along the beam direction (Fig. 1). It is made of copper (cp), aluminum (al) or aluminum with holes (alh), depending on the desired mixed-field intensity and composition;
13 rack test locations in lateral and downstream positions with respect to the beam direction, 3 ÷ 5 m far from the target (Fig. 1);
a shielding between the target and the lateral positions, created by using 4 movable blocks of concrete (C) and iron (I) (Fig. 1).
The 24 GeV/c proton beam enters the CHARM facility and impinges on the target described above. From the interaction between the two, a shower of particles takes place. This shower can further interact with the shielding (if any), possibly generating other secondaries and consequently producing (or stopping) other particles: this is the way we create and tune the mixed-field at CHARM.
In particular, in this paper, two target-shielding layouts are examined: copper target and no shielding (movable plates in OFF position), namely cp_0000, and copper target and concrete–iron–iron–concrete shielding, i.e. cp_CIIC. Other mixed-fields can be produced using different shielding configurations and target materials.
FLUKA Monte Carlo simulations have already been performed to characterize the different radiation fields: the knowledge of the field at any rack test locations in terms of particles spectra is the first, preliminary step to run the dedicated simulations on the RadFET (first simulation step).
Fig. 1 CHARM test facility model, top view. |
2.2 FLUKA simulations settings to calculate the mixed-field at CHARM
A FLUKA model of CHARM was built in the recent past to simulate the radiation field inside the test facility and retrieve all the most relevant quantities for radiation to electronics testing.
The NEWDEFA1 FLUKA default was used as a first approximation to reduce the computational time, given the large geometry model of the facility: the higher the thresholds for production and transport of particles, the lower the computational time.
For the same reasons, a 20 cm side air cube was chosen as sensitive volume: to score the physical quantities of interest (dose, energy, fluences, etc.). Scoring, for instance, dose inside a the gate oxide (GO) of the RadFET – whose characteristic dimensions are several order of magnitude lower than the CHARM's size –, would have led to a disproportional increase of the CPU time (a larger amount of primaries to get the same statistics).
An example of the characterization of the mixed-field at CHARM in terms of simulated particle energy spectra (lethargy), for the cp_0000 and cp_CIIC configurations is presented – for the position 1 – in Figure 2. It is clear how the field is populated by secondary electrons, positrons, muons (μ±), photons, neutrons and charged hadrons (protons, kaons (κ±) and pions (π±)). Those spectra have been obtained from FLUKA, making the 24 GeV/c proton beam interact with a copper target and simulating the entire CHARM geometry, reproducing the two different shielding layouts.
From Figure 2 it is clear, for instance, that the lethargy2 is overall lower for all the particles of the mixed field in the cp_CIIC configuration. The CIIC walls are stopping a certain amount of particles of the shower, preventing them from reaching the rack position 1.
Fig. 2 Mixed-field: simulated spectra (lethargy), position 1, cp_0000 and cp_CIIC configurations. |
2.3 Experimental measurements and simulated dose: benchmark
Dose at CHARM is experimentally detected by an integrated dosimetry system, called RadMon (Radiation Monitoring system) [9,10]. The RadMon executes an electronic readout of the threshold voltage of the RadFETs and sends the voltage value to the CERN database. The conversion from voltage (Vgs) to dose (Gy) is post-processed by means of a calibration curve, obtained for each RadFET batch in a Co60 source. The experimental results presented in this paragraph refer to the data measured during 2015 run period of CHARM, for the copper target and no shielding configuration, measured by 100 nm p-channel RadFET mounted on RadMon. The experimental doses in Table 1 are averages of the calibration factors retrieved during different runs of 2015 plus minus total associated errors, which take into account primary proton beam intensity error, RadFET error and standard deviation with respect to the average value [6].
It is worth noticing that the previous FLUKA simulations results (obtained by scoring the dose in a cubic volume of air, 20 cm side, placed in the different test positions, considering the beam-target interaction and the whole CHARM geometry and setting energy thresholds for transport and production of secondaries at 1 MeV) in Table 1 are systematically overestimating the measured doses of 25% and 43%.
Figure 3 shows a plot of the trend of dose simulated in different positions in the test area. The errors associated to the simulated values (as also clear from Tab. 1) are low if compared to the experimental ones. They take into account only statistical uncertainties and no systematic errors, like in the case of measured doses.
Experimental and 1st step simulated doses in 10−14 Gy/pot, in different positions at CHARM, cp_0000 configuration.
Fig. 3 Experimental and simulated doses at different test positions, cp_0000 target-shielding layout. |
3 The RadFET model and the second step simulations: benchmarks
We want now to investigate the discrepancies between the performed FLUKA simulations and the measurements detected by the RadMON (position 1, cp_0000 layout) to evaluate if it is possible to better simulate the radiation dose levels at CHARM. If we can improve the simulations settings, in order to obtain doses that are more realistic and in line with experimental measurements, we could use our calculations to predict the radiation field at each position. And this can in principle be done for each target-shielding layout at CHARM, even out of the 14 ordinary predefined positions.
In order to improve the reliability of our MC simulations and the agreement with respect to the experimental data, we decide to implement a more realistic second step simulation acting on:
geometry and materials of the sensitive volume;
energy thresholds for production and transport of particles;
energy cutoffs for input spectra.
We want to conduct different parametric analyses to find out which are the most accurate FLUKA simulation setting.
In the following paragraph, the first step dose calculated in correspondence of the position 1, for the copper target and no shielding configuration (Sects. 2.1 and 2.2), is taken as reference and its value is 1.101 × 10−14 ± 0.8% Gy/pot.
Before going through the details of the parametric simulations, we need to put in relation the two steps: if the geometry and the size of the sensitive volume are going to change, also the overall geometry and scale of the problem have to change accordingly.
3.1 Link between the first and second simulations step
As already specified (Sect. 2.1), the spectra of particles characterizing the CHARM mixed-field had already been retrieved in the first step simulation (Sect. 2.2). Those spectra are now available and can be used as input to perform the second step simulations: the RadFET will be irradiated by a rectangular beam carrying information about the mixed-field in a given position and for a specific target-shielding layout. The beam has the same surface as the RadFET crossing section (1 mm2). In order to give an example, if we want to simulate the dose response of the RadFET, we want to irradiate the actual geometry of the MOSFET using a beam that is carrying with itself the spectra of protons, electrons, positrons, muons, etc., calculated in the first step simulation, in desired position and target-shielding configuration.
The approximation of the radiation field at CHARM as a mono-directional beam is realistic as the real mixed-field is in good approximation radial. The particle shower originating from the target (and possibly interacting with the shielding, if present) reaches all the rack test positions moving almost along radial directions.
3.2 The RadFET geometry and materials: their role in the second step
As shown in Figure 4, where the yz (z is assumed to be the beam direction) and the xy views are shown, the RadFET is mainly composed of a 250 μm kovar lid, a vacuum region of 250 μm and a 500 μm Silicon substrate; the crossing plane (xy) surface is 1 × 1 mm2. A 1.4 μm Silicon die is then located on the Silicon substrate: the SiO2 gate oxides we are interested to study are deposited on this thin layer (Fig. 4). The dioxides, 4 in total in our RadFET (named ox1, ox2, ox3, ox4), can be 100 or 400 nm thick and both the configurations are studied in this paper. For our simulations we took into account just the ox1 Gate Oxide, having dimensions 300 μm × 50 μm × 400 nm (or 100 nm), centered in x = 0.000 cm and y = −0.0135 cm, with respect to the RadFET center.
The second step approach has been then validated by reproducing the first step configuration (air cube of 20 cm side) for position 1 in copper target and shielding OFF configuration. We got a deposited dose in the air rack of 1.101 × 10−14 ± 0.7% Gy/pot. This is in perfect agreement with the dose calculated in the 1st step, obtained from a more realistic and complex geometry model of the entire test facility, with a more realistic mixed-field. We are therefore confident that the approximation we introduced to reduce the CPU time, given the microscopic dimensions of our sensitive volume (mixed-field given through a directional beam), is good enough to faithfully reproduce the 1st step.
After this validation of the model, we decide to calculate dose in the 6 × 10−6 mm3 gate oxide (SiO2) of the RadFET, in kovar lid configuration, using the same energy thresholds as before. The simulated dose in this case is 0.895 × 10−14 ± 2.5% Gy/pot, showing a reduction of 19%, if compared to the dose in the air volume.
As clear in Table 2, the deposited dose shows a further decrease if we use the PRECISIO physics settings plus 1 keV energy thresholds: in this case the dose is 0.757 × 10−14 ± 2.0% Gy/pot, 15% lower than the one obtained with the NEWDEFA settings and thresholds at 1 MeV and 31% smaller than what was calculated on the air volume, using the same settings as in the 1st step simulation.
This evidently indicates that if we set a more accurate simulation in terms of geometry, materials and energy thresholds, the deposited dose shows a significant decrease with respect to that obtained in the 1st step simulation.
Fig. 4 RadFET geometry and materials: FLUKA model. |
Dose in 10−14 Gy/pot deposited within the air rack and the RadFET gate oxide in SiO2, using 1 MeV (with NEDEFA defaults) and 1 keV (using PRECISIO defaults) energy thresholds.
3.3 Update of the energy thresholds
As already discussed, the use of the NEWDEFA default in the first step simulation was crucial to reduce the CPU time, given the large geometry of the CHARM test facility. On the other hand, considering that the range of secondary electrons in air is 376 cm [11] and that the characteristic dimension of the air scoring volume we are using is 20 cm, lowering thresholds represents an improvement of the accuracy of our simulations (first row in Tab. 2). It is important to point out, in fact, that in the case of 1 MeV thresholds, we are clearly overestimating the energy deposited, introducing an artifact: all the particles having energies lower than 1 MeV are treated by the FLUKA Monte Carlo algorithm as depositing their energy on spot, without producing other particles and being transported.
Following the previous considerations, the second step simulations can be then mostly conducted using the PRECISIO default, mainly characterized by particles transport thresholds set at 100 keV (except for neutrons, 10−5 eV) and delta ray production threshold at 100 keV. In addition, by using the EMF-CUT and DELTARAY cards, thresholds for transport and production of secondary electrons, positrons and photons and delta ray generated by muons and charged hadrons will be set at 1 MeV, 100 keV, 10 keV and 1 keV. The PART-THR card is instead employed to lower energy transport cut-offs for protons, pions, muons, kaons and α particles to 1 keV.
Given the 15% difference in the dose deposited in the RadFET gate oxide, going from 1 MeV to 1 keV (last row in Tab. 2), it is worth investigating now in which thresholds energy range this main change occurs. The results of this parametric analysis can be found in Table 3.
The main change in dose occurs between 1 MeV and 100 keV, showing a decrease of 14%. Going down to 10 keV and 1 keV does not considerably change the scenario, and the differences are in the range of the statistical uncertainties. This behavior can be explained considering the range of electrons in SiO2: 1 MeV electrons have a range of 1.96 mm, 100 keV electrons have a range of 66.15 μm, while 10 keV electrons have a range of 1.23 μm. By comparing these ranges of secondary electrons with the characteristic lateral dimensions of the gate oxide of the RadFET (300 μm × 50 μm), it is evident that thresholds lower than 100 keV are the most appropriate to simulate and faithfully reproduce the energy deposition in the gate oxide of the RadFET.
In this specific case, since the beam surface has the same size as the crossing surface of the RadFET and the ox1 sensitive volume is completely irradiated by the beam itself, we have a good statistics using a reasonable number of primaries: we decide to lower the thresholds down to 1 keV to have most precise simulated doses, without losing big amounts of CPU time.
Parametric analysis changing the energy thresholds of the simulations (PRECISIO defaults): dose in the SiO2 GO of the RadFET, in 10−14 Gy/pot.
3.4 Input spectra cutoffs and oxides thickness
After evaluating the dose dependency on materials, scoring volume sizes and energy thresholds, we decide now to study what is the impact of the oxide thickness and how input spectra cutoffs could affect the final dose.
We simulate RadFET having 100 nm and 400 nm thick gate oxides. We reproduce, on the other hand, input spectra characterizing the mixed-field, through a dedicated and independent first step simulation, by cutting them at 1 MeV and 100 keV, to see how lower energy particles of the spectra could affect the final deposited dose. Going down to energies lower than 100 keV is counterproductive: the lower the thresholds, the lower the ranges and that particles can be stopped by the kovar lid of the RadFET. From Table 4, the oxide thickness does not have a large impact on the simulated deposited dose. On the other hand, using spectra cut at 1 MeV or 100 keV makes a difference that is of the order of 10%.
Effect of gate oxide thickness and input spectra thresholds on the RadFET dose response: simulated dose in 10−14 Gy/pot (PRECISIO defaults set).
4 Second step simulations: the cp_0000 target-shielding configuration
The analyses conducted in Section 3 led us to identify the following simulation settings as the most suitable for our second step simulations:
400 nm thick oxide;
PRECISIO default in FLUKA;
energy thresholds at 1 keV;
input spectra cutoffs at 100 keV.
4.1 Doses simulated and measured in different positions at CHARM
We set the RadFET to positions 3, 5, 7, 10 and 13, by using as input the spectra of the 1st step simulations, calculated for the new locations of interest, as described in Section 2.2.
Figure 1 shows how the new considered locations are differently exposed to the mixed-field, mostly because of geometric reasons: we expect the position 13 to be actually impacted by the highest beam intensity, since directly exposed to the flux and almost in the beam direction. In addition, we want to compare simulated doses to experimental data (Sect. 2.3), to show the improvements we got by performing a more accurate second step simulation.
Table 5 and Figure 5 show how the difference between experimental data and simulated values is now lower if compared to that obtained in Sections 2.2 and 2.3. The mismatch measured-simulated is 5 ÷ 25% – compatible with experimental uncertainties instead of 25 ÷ 43%, showing a substantial reduction (20 ÷ 25%) of the simulated dose (Tab. 6), in the direction of experimental data.
It is then evident that the new simulation allowed us to obtain most accurate simulated values, in better agreement with the measured doses. This is also clear from Figure 5, where experimental, 1st and 2nd step doses are plotted: the second step simulation has represented an important improvement in simulating the radiation field at CHARM in the copper target and shielding OFF configuration.
Fig. 5 Experimental and simulated doses – 1st and 2nd step simulations – at different test positions, cp_0000 configuration. |
Experimental and 2nd step simulated doses in 10−14 Gy/pot, in different positions at CHARM.
1st and 2nd simulated doses in 10−14 Gy/pot, in different positions at CHARM.
4.2 Contribution of each mixed-field's particles to the total dose
The simulations results of the single particles (of the mixed-field) contribution to the total dose are now presented. The dose coming from all the particles of the mixed-field is going to be tabulated for the positions 1, 3, 10 and 13; particular attention will be given to the weight of each particle over the total dose.
As shown in Figures 6 and 7, downstream positions (10 ÷ 13) are the most exposed to the particle shower developing after the beam-target interaction. Speaking in terms of doses deposited by the different particles of the mixed-field, the energies deposited by charged hadrons, electrons and positrons, muons increase going from location 1–13 (Fig. 7). On the other hand, doses coming from neutrons and photons are almost constant. The explanation has to be found in the topography of our radiation field: fluences of charged hardons, electrons and positrons an muons are in fact characterized by a forward directionality, while, on the contrary, photons and neutrons fluences are more isotropic.
Considering the relative contribution of each particle of the field over the total dose (Fig. 6), it is interesting to note how, going from lateral to downstream positions, from 1 to 13:
charged hadrons (protons, pions, kaons) contribution rises from 34% to 48%;
electrons and positrons contribution rises from 30% to 40%;
muons doses are stable around the 3–4%, representing an almost negligible portion of the total;
photons doses decrease from 24% to 7% in terms of contribution to the total;
neutrons contribution goes down from 8.5% to 2.5% in terms of contribution to the total.
As already said, this is a specular way to see how going from lateral to upstream positions, the contributions of the primary particles of the mixed-field to the total dose change and in particular charged hadrons, electrons and positrons become dominant, while photons and neutrons weights lose relevance.
Fig. 6 Particles contribution to the total dose: relative contribution. |
Fig. 7 Particles contribution to the total dose: dose values. |
5 cp_CIIC target-shielding configuration: the thermal neutrons issue
In this paragraph we will use the same simulation settings employed in the frame of the cp_0000 target-shielding configuration, for the cp_CIIC layout.
As for the previous case, we will irradiate the RadFET with a beam carrying information about the different spectra of the mixed-field, starting from position 1.
5.1 Doses simulated and measured in position 1
We want to compare the two-step simulations results. As visible in Table 7, where doses for position 1 and cp_CIIC are reported, the mismatch between first and second step doses is now 52%. This could mean that probably something is missing in our model.
Given the unexpected discrepancy, we decide to evaluate the single contributions to the total dose, due to all the particles populating the CHARM mixed-field, in cp_CIIC configuration. In this regard, Figure 8 shows a comparison between the two step in the air cube of 20 cm side (taken as a reference) and the RadFET gate oxide, using the same settings as in Section 3.1. It is worth noticing that the doses deposited by charged hadrons, muons and electrons + positrons + photons are perfectly comparable in both cases (air volume and RadFET). What really changes is the dose contribution due to neutrons: the dose simulated in the large air volume is a factor 5.3 higher than what scored in the RadFET.
2nd and 1st step simulated doses in 10−14 Gy/pot, in position 1, cp_CIIC configuration.
Fig. 8 Particles contribution to the total dose: dose values. |
5.2 Splitting neutrons spectrum: contribution at different energies
For the reasons described in the previous paragraph, we decide to split the spectrum of neutrons in different energy ranges, to further investigate where this mismatch could be originated.
From Table 8, it is clear that the discrepancy we got is mainly due to thermal neutrons, which are depositing in the RadFET gate oxide a dose that is a factor 50 lower than expected (dose simulated in the air volume and measurements). This is, anyway, something in line with expectations: thermal neutrons undergo np nuclear interactions with the nuclei of Nitrogen atoms present in air and these interactions give birth to protons at 300 ÷ 400 keV that are actually depositing energy in the air cube volume.
On the other hand, this is evidently not the case if we consider the dose simulated in the RadFET gate oxide: before reaching the oxide, thermal neutrons have interacted with the Kovar (Ni-Co) lid (possibly developing a further particle shower) and finally deposited energy in SiO2, without interacting with any air molecule.
This is clearly only partially explaining the discrepancy we found and further investigations are already ongoing, laying the basis for the next step of research. In the next future, in fact, we want to build a most accurate model of the RadFET in terms of materials, thinking, for instance, that the doping could be a potential source of this discrepancy. In parallel, new test runs at CHARM will give us more information and most reliable measurements about the dose absorbed by RadFETs in the cp_CIIC configuration. Also a test campaign to measure the RadFET sensitivity to low energy neutrons will take place in the early 2017.
Contributions to the total dose deposited in the air cube and the RadFET GO, in 10−14 Gy/pot, in position 1, due to neutrons at different energies.
6 Summary and conclusions
By implementing the real geometry of the RadFET in FLUKA, using most accurate physics settings and input spectra from the first simulation step, we are able to simulate the RadFET dose response in the CHARM mixed-field – copper target and shielding OFF configuration – with a high accuracy. Through the second simulations step, we got a reduction of the mismatch between measured and simulated doses of 20 ÷ 25%. Equivalently, simulated doses are definitely closer to the experimental data (5 ÷ 25% difference between the two), than what had already been done with the previous calculations, which shown a difference of 25 ÷ 45% between simulated and measured values.
On the other hand, the same simulation settings do not seem to be entirely applicable when simulating the dose response of the RadFET under the mixed-field generated in copper target and concrete–iron–iron–concrete shielding configuration. From the simulations we conducted, this is apparently due to the thermal neutrons component of the radiation field at CHARM (dominant in the shielded case and evidently negligible in the cp_0000 configuration) and requires further investigation both from the experimental and simulation point of view. In this regards, this is the starting point for the next future research: the final objective is to find unambiguous simulation settings to be used under different target-shielding configurations at CHARM and confirmed by experimental measurements.
References
- P. Beck et al., Investigation on photon energy response of RadFET using Monte Carlo simulations, IEEE Trans. Nucl. Sci. 54, 1151 (2007) [CrossRef] [Google Scholar]
- M. Wind et al., Investigation of the energy response of RadFET for high energy photons, electrons, protons, and neutrons, IEEE Trans. Nucl. Sci. 56, 3387 (2009) [CrossRef] [Google Scholar]
- J. Mekki et al., Mixed particle influence on RadFET responses using Co-60 calibration, IEEE Trans. Nucl. Sci. 60, 2435 (2013) [CrossRef] [Google Scholar]
- A. Holmes-Sieldle, L. Adams, Handbook of radiation effects, 2nd ed. (Oxford University Press, 2002) [Google Scholar]
- J. Mekki et al., in A mixed field facility at CERN for radiation test: CHARM, RADiation and its effects on components and systems (RADECS), 14–18 September (2015) [Google Scholar]
- S. Bonaldo, CHARM Una nuova facility del CERN per test di elettronica con spettri misti di radiazione ionizzante, MSc thesis, University of Padova, 2016 [Google Scholar]
- A. Ferrari et al., FLUKA: a multi-particle transport code, CERN-2005-10, INFN/TC05/11, SLAC-R-773 (2005) [Google Scholar]
- T.T. Böhlen et al., The FLUKA code: developments and challenges for high energy and medical applications, Nucl. Data Sheets 120, 21 (2014) [Google Scholar]
- G. Spiezia et al., The LHC accelerator radiation monitoring system – RadMon (2011) [Google Scholar]
- G. Spiezia et al., A new RadMon version for the LHC and its injection lines, IEEE Trans. Nucl. Sci. 61, 3424 (2014) [CrossRef] [Google Scholar]
- M.J. Berger et al., ESTAR, PSTAR, and ASTAR: computer programs for calculating stopping-power and range tables for electrons, protons, and helium ions (version 1.21), http://physics.nist.gov/Star [Google Scholar]
Cite this article as: Matteo Marzo, Stefano Bonaldo, Markus Brugger, Salvatore Danzeca, Ruben Garcia Alia, Angelo Infantino, Adam Thornton, RadFET dose response in the CHARM mixed-field: FLUKA MC simulations, EPJ Nuclear Sci. Technol. 3, 24 (2017)
All Tables
Experimental and 1st step simulated doses in 10−14 Gy/pot, in different positions at CHARM, cp_0000 configuration.
Dose in 10−14 Gy/pot deposited within the air rack and the RadFET gate oxide in SiO2, using 1 MeV (with NEDEFA defaults) and 1 keV (using PRECISIO defaults) energy thresholds.
Parametric analysis changing the energy thresholds of the simulations (PRECISIO defaults): dose in the SiO2 GO of the RadFET, in 10−14 Gy/pot.
Effect of gate oxide thickness and input spectra thresholds on the RadFET dose response: simulated dose in 10−14 Gy/pot (PRECISIO defaults set).
Experimental and 2nd step simulated doses in 10−14 Gy/pot, in different positions at CHARM.
2nd and 1st step simulated doses in 10−14 Gy/pot, in position 1, cp_CIIC configuration.
Contributions to the total dose deposited in the air cube and the RadFET GO, in 10−14 Gy/pot, in position 1, due to neutrons at different energies.
All Figures
Fig. 1 CHARM test facility model, top view. |
|
In the text |
Fig. 2 Mixed-field: simulated spectra (lethargy), position 1, cp_0000 and cp_CIIC configurations. |
|
In the text |
Fig. 3 Experimental and simulated doses at different test positions, cp_0000 target-shielding layout. |
|
In the text |
Fig. 4 RadFET geometry and materials: FLUKA model. |
|
In the text |
Fig. 5 Experimental and simulated doses – 1st and 2nd step simulations – at different test positions, cp_0000 configuration. |
|
In the text |
Fig. 6 Particles contribution to the total dose: relative contribution. |
|
In the text |
Fig. 7 Particles contribution to the total dose: dose values. |
|