Dosimetric properties of the most commonly used dosimeters in medical physics and personal dosimetry [8,9,1223].

Mat. Zeff Doping Dosimetera [6Li] Linearity rangeb Repeatability Uniformity (batch, 1σ) Fading Relative sensitivity Applications

                  /TLD-100 nth/γ  
natLiF 8.2 Mg,Ti TLD-100 7.5% [9]
7.36% [12]
10−5–10 Gy [8,13] <2% [13] 7.5% [13] 5% y−1 [13]
5–10% y−1 [8]
1 [13]   Medical physics
      MTS-N   5×10−5–5 Gy [14] <2% [14] <5% [14] <5% an−1 [14] 1.5 [14]  
      GR-100   10−7–12 Gy [15]     <5% y−1 [15]    
    Mg,Cu,P TLD-100H 7.36% [12] 10−6–10 Gy [13]
10−6–20 Gy [8]
    Negl. [13]
3% y−1 [8]
15 [13]   γ, β dosimetry
personal, environment
      MCP-N   10−7–10 Gy [14] <2% [14] <5% [14] <5% y−1 [14] 40 [14]  
      MCP-Ns   10−5–10 Gy [14] <2% [14]   <5% y−1 [14] 1 [14]  
      GR-200   10−7–12 Gy [15] <2% [15]   <5% y−1 [15] 65 [15]  
6LiF 8.2 Mg,Ti TLD-600 95.6% [9] 10−5–10 Gy [13]
10−5–20 Gy [8]
    5% y−1 [13]
Negl. [8]
1 [13] 350 [16] Medical physics
      MTS-6   5×10−5–5 Gy [14] <2% [14] <5% [14] <5% y−1 [14] 1.5 [14]  
  
    Mg,Cu,P TLD-600H 95.12% [12] 10−6–10 Gy [13]
10−6–20 Gy [8]
    negl. [8,13] 15 [13] 220 [16,17] Neutron dosimetry personal, environment
      MCP-6   10−7–10 Gy [14] <2% [14] <5% [14] <5% y−1 [14] 40 [14] 127 [16]
      MCP-6s   10−5–10 Gy [14] <2% [14]   <5% y−1 [14] 1 [14] 450 [17]
      GR-206   10−7–12 Gy [15] <2% [15]   <5% y−1 [15]   192 [16,17]
a

TLD-100 to 900 are synthesized by Harshaw (Thermo Fisher Scientific Inc.), MTS and MCP TLDs by TLD Poland, GR TLDs by SDDML (Solid Dosimetric Detector and Method Laboratory), PTL-717 TLDs by Desmarquest-CEC, and nanoDots OSLDs by Landauer.

b

The luminescent responses of the dosimeters are independent of the dose rate over the considered linearity ranges (up to1000 MGy s−1 for Harshaw TLDs).