Open Access

Table 1

PTB calibration for different energies in the 1–20 MeV range. Flux rates correspond to a distance of 1 m from the source. Ti(T) stands for tritiated titanium. The energy spread is represented by the FWHM. The relative contribution of neutrons scattered in the target is indicated by Φsc/Φ.

Reaction <En> (MeV) (MeV) Target Flux (/cm2/s) Φsc/Φ (%)
2H(d,n)3He 5.0 0.2 D2-gas 5.2·103 <1.0
2H(d,n)3He 8.0 0.2 D2-gas 1.9∙104 <1.0
3H(d,n)4He 14.8 0.4 Ti(T) 1.3∙104 3.0

Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.

Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.